NanoPhotometer™ Pearl Design Edition - Small. Fast. Accurate.
NanoPhotometer™ Pearl Design Edition UV/VIS spectrophotometer for Nano Volume and standard cuvette applications with built-in computer, touch pad, thermal paper printer, USB port for data transfer, NanoPhotometer™ Pearl Software Package, cables and manual. (Software Sold Separately)
The NanoPhotometer™ Pearl combines ultra low (0.3µl) and standard volume cuvette applications for improved concentration measurement and purity analysis in one UV/Vis spectrophotometer The NanoPhotometer™ Pearl is a maintenance free and stand alone device with flexible data output options, bringing portability, convenience, and speed to UV/Vis spectrophotometry. Applications are limitless with full scanning (200-950nm), superior sample stability, and programmable methods to suit your needs. The NanoPhotometer™ Pearl Design Edition comes with lid factor 10 and 50.
Product Features:
Average Detection Range for Nucleic Acids and Proteins
Product Specifications:
SPECTROMODULE
|
190 - 1,100 nm |
Wavelength scan range |
200 - 950 nm |
System start up time |
Less than 5 seconds, no warm up necessary |
Measure time for full scan range |
3.5 seconds |
Wavelength reproducibility |
< ± 0.2 nm |
Wavelength accuracy |
± 2nm |
Bandwidth |
Better than 5nm |
Stray light |
<0.5% at 220 nm using NaI and 340 nm using NaNO2 |
Photometric range |
-0.3 - 2.499 A 0-199% T |
Detection Range |
dsDNA: 0.5 ng / µl to 125 ng / µl, BSA: 0.02 mg / ml to 3.6 mg / ml |
Absorbance reproducibility |
±0.003 A (0 to 0.5 A), ±0.007 A (0.5-1.0 A) @ 260 nm |
Absorbance accuracy |
±0.005 A or ±1% of the reading, whichever is the greater |
Zero stability |
±0.003 A/hour after 20 min warm up @ 340 nm |
Noise |
0.002 A rms at 0 A @ 260 nm |
Optical arrangement |
Dual channel Czerny Turner with flat grating, 1024 pixel CCD array, concave mirrors |
Lamp |
Xenon flash lamp |
Lifetime |
10⁹ flashes, up to 10 years |
Cell types |
15 mm centre height (z-height), outside dimension 12.5 mm x 12.5 mm |
SMALL VOLUME OPTION
|
dsDNA: 2 ng/µl to 18,750 ng/µl, BSA: 0.08 mg/ml to 543 mg/ml |
Absorbance range |
0.01 - 1.5 (10 mm equivalent: 0.05 - 375) |
Minimum sample size |
0.3 µl |
Path lengths |
0.04 mm, 0.1 mm, 0.2 mm, 1 mm and 2 mm |
Virtual dilution factors |
5, 10, 50, 100 and 250 |
OTHER TECHNICAL DATA
|
capacity for eight 10 mm cells |
Photometric mode |
Abs, %T, concentration, scan, ratio, multi wavelength, kinetics in Abs x factor / min |
Method storage |
Up to 81 methods in user methods |
Built-in methods |
Nucleic acid, microarray (labeling efficiency), protein and cell density |
Display formats |
320 x 240 pixels |
Size |
140 mm x 275 mm x 380 mm |
Weight |
< 4.5 kg |
Operating voltage |
90-250 V, 50/60 Hz, Max 30 VA |
Input / Output ports |
SD Memory Card, USB or Bluetooth for connection to a PC for direct data download, printout and data storage |
Performance verification |
Auto diagnostics when switched on |
Warranty |
1 year |